Recent Changes - Search:

radlab home

radlab support

generic HFDR

principles
hardware
software
production
documents
pictures

know-how

projects

CNMI-Guam
IFREMER
ISMER
PACIOOS
UABC
UHHilo
UP-MSI
WHOI

old projects

MEC
OGS
UAF

wiki instructions

.

PA112

PA 1-12 Updated (03/21/2016)

Test Frequency : 8.25 MHz TOMCO serial number 6163 0.1-30 MHz
WERLATONE serial number 96389 1-30 MHz
FILTER (9MHz) serial number 0002 (*doug williams)

Settings

  • AUTO DHCP - ok
  • FRONT LED - ok; when no load, LED OFF - ok
  • default dB in = 0dB

Power Supply Measurements

AC voltage setting: 115
DC voltage (V): 26.5 (should be about 24.5 V under load, adjust otherwise)
Ammeter reading (A): 1.8
UPS power load w/ 2dBm input(W): N/A

Signal Measurements

  • RF power meter: Mini Circuits PWR-SEN-6GHS USB power sensor
  • internal attenuator set at 0 dBm
  • power meter 50 dBm offset applied for measurements at rear panel connectors
  • 10 dB attenuators used at FWD and RFL outputs to achieve 50 dB atten of actual signal
LocationMeter Reading (dBm)Rabbit Reading (dBm)Comments
IN connector1.861.3 
FWD connector   
RFL connector   
OUT connector47.8847.5 

Temperature Measurements

  • temperature meter: DM6802B Digital Thermometer
LocationMeter Reading (deg C)Rabbit Reading (deg C)Comments 
BTC40.642.2 
MCI25.524.9 
AMB30.429.9 

old Notes

Spectrum Analyzer Images

Input Signal

At FWD connector

At RFL connector

At OUT connector

Edit - History - Print - Recent Changes - Search
Page last modified on March 22, 2016, at 12:57 am