PA 1-09 Updated 03/21/2016

Test Frequency : 8.25 MHz TOMCO serial number 6162 wrong serial 0.1-30 MHz
WERLATONE serial number 96391 1-30 MHz
FILTER (9MHz) serial number 0001 (*doug williams)
Settings

  • AUTO DHCP - ok
  • FRONT LED - ok; when no load, LED OFF - ok
  • default dB in = 1 dBm

Power Supply Measurements

AC voltage setting: 115 V
DC voltage (V): 26.5
Ammeter reading (A): 1.7 Amps
UPS power load w/ 2dBm input(W): Not Available

Signal Measurements

  • RF power meter: Mini Circuits PWR-SEN-6GHS USB power sensor
  • internal attenuator set at 0 dBm
  • power meter 50 dBm offset applied for measurements at rear panel connectors
  • 10 dB attenuators used at FWD and RFL outputs to achieve 50 dB atten of actual signal
LocationMeter Reading (dBm)Rabbit Reading (dBm)Comments
IN connector1.861.5 
FWD connector------ 
RFL connector------ 
OUT connector47.1946.8 

Temperature Measurements

  • temperature meter: DM6802B Digital Thermometer
LocationMeter Reading (deg C)Rabbit Reading (deg C)Comments 
BTC29.233.8Meter Reading TOMCO CASE
MCI22.718.5Meter Reading on Power Sensor3
AMB23.225.9Meter reading on back compartment

(old notes)

Spectrum Analyzer Images

Input Signal

At FWD connector

At RFL connector

At OUT connector