Notes
- TOMCO serial number 6168
- this unit has been configured for 12 MHz
- shipped to UABC (Flores)
Power Supply Measurements
AC voltage setting: 115V
DC voltage (V): 24.0
Ammeter reading (A): 1.1
UPS power load w/ 2dBm input(W):
Signal Measurements
- RF power meter: Mini Circuits PWR-SEN-6G+ USB power sensor
- internal attenuator set at 0 dBm
- power meter 50 dBm offset applied for measurements at Tomco output and rear panel connectors
- calibration coefficients entered into GUI to correct Rabbit readings (IN,FWD,RFL): 1.2, 0.6, 0.84
Location | Meter Reading (dBm) | Rabbit Reading (dBm) | Comments | |
IN connector | -0.1 | -0.1 | |
FWD connector | +44.19 | +44.3 | |
RFL connector | ~ +13.4 | +13.4 | |
OUT connector | +44.22 | -- | |
Temperature Measurements
- temperature meter: DM6802B Digital Thermometer
Location | Meter Reading (deg C) | Rabbit Reading (deg C) | Comments | |
BTC | * | 30.2 | |
MCI | ~23 | **** | wire loose, will be fixed after test |
AMB | 25 | 26.6 | |
Spectrum Analyzer Images
Input Signal
At FWD connector
At RFL connector
At OUT connector