Notes
- TOMCO serial number 6164
- LPF 12 MHz
- Shipped to UABC (Durazo)
Power Supply Measurements
AC voltage setting: 115 V
DC voltage (V): 24.0
Ammeter reading (A): 1.3
UPS power load w/ 2dBm input(W): ???
Signal Measurements
- RF power meter: Mini Circuits PWR-SEN-6G+ USB power sensor
- internal attenuator set at 0 dBm
- with input at +2.02 dBm, output +1.77 dBm through LPF and DC to output connector
- power meter 50 dBm offset applied for measurements at Tomco output and rear panel connectors
Location | Meter Reading (dBm) | Rabbit Reading (dBm) | Comments | |
IN connector | +2.02 | -0.6 | |
FWD connector | +46.01 | +45.5 | |
RFL connector | +5.0 +/- 0.5 | +5.1 | |
OUT connector | +46.02 | x | |
Temperature Measurements
- temperature meter: DM6802B Digital Thermometer
Location | Meter Reading (deg C) | Rabbit Reading (deg C) | Comments | |
BTC | 24 | 33.8 | |
MCI | 27 | 23.3 | |
AMB | 24 | -3.0* | *no thermistor installed |
Spectrum Analyzer Images
Input Signal
At FWD connector
At RFL connector
At OUT connector